+86 027-65279286  marketing@blcd.cc

Insulation Oil Test Equipment

BL-330 Insulating Oil Dielectric Loss Tester

  • Product Class: Insulation Oil Test Equipment
  • Phone: +86 135 5410 7427
  • Tel: +86 027-65279286
  • Email: marketing@blcd.cc
  • Feedback

Product Introduction

    In the preventive insulation test of power equipment, it is required to measure the insulation oil parameters of power equipment periodically. The measurement of dielectric loss and resistivity of insulating oil is one of the most important. For a long time, most of them adopt the bridge method for measurement, which is complicated in operation and the measurement accuracy is affected by many factors, thus leading to a large measurement error. With the rapid development of electronic technology and the power industry for small volume, light weight, easy to operate, rapid measurement, high precision measurement instrument requirements, I refer to domestic and foreign related instruments developed in the domestic leading insulation oil dielectric loss factor and dc resistivity tester. The instrument is designed and manufactured according to gb5654-2007 and related standards. It is controlled by microcomputer, easy to use, high measurement accuracy, high test efficiency and greatly reduces labor intensity.

Product Parameters

Test voltage range         0~2000V AC,0~500V DC
Test Capacitance Range         50pF~200pF
Test Temperature Range         at room temperature~125℃
Dielectric Loss Test Range         0.00001~1
Accuracy of Measurement         ±(value×0.5%+0.0001)
Relative Permittivity         ±(value×0.5%+0.1)
Resistivity Resolution         0.001MΩ·m
Resistivity Measurement Range         2.5MΩ·m~20TΩ·m
Power         500W
Power Supply Voltage         AC220V±22V
Environment Temperature         0℃~+40℃
Relative Humidity         ≤75%RH
Overall Dimensions         470×430×380mm3
Weight         23kg

Product Characteristics

    ● The structure of the instrument is integrated with functions of oil cup, heating, temperature control and pressure regulation.

    ● The use of large screen color LCD, Chinese thermal printing, Chinese menu, simple operation.

    ● Empty cup automatic calibration.

    ● With over voltage, over current, temperature limit protection function.

    ● Medium frequency induction heating electrode cup, short time uniform heating.

    ● Directly measure the temperature through the probe placed in the measuring electrode cup.

    ● It includes sine wave generator and digital voltage regulation to produce standard 50Hz high-power test power.

LATEST NEWS

CONTACT US

QQ:2371515083

Phone:+86 135 5410 7427

Tel:+86 027-65279286

Email:marketing@blcd.cc

Add:Building 207, Magic Square Building, No.303 Guanggu Avenue, East lake high tech Development Zone, Wuhan, China