AC/DC High Test Set

BL-540A Integrated DC Hipot Tester

  • Product Class: AC/DC High Test Set
  • Phone: +86 135 5410 7427
  • Tel: +86 027-65279286
  • Email: marketing@blcd.cc
  • Feedback

Product Introduction

    BL-540A Integrated DC Hipot Tester is a new generation of portable high-voltage detection instrument designed and manufactured according to China Power Industry Standard DL/T848.1-2004 "High-voltage test equipment general technical conditions Part 1: DC Integrated dc high voltage generator". Electrical insulation strength and leakage current of power devices.

    The Integrated dc high voltage generator uses a high-frequency voltage doubling circuit, which uses advanced PWM pulse width modulation technology and voltage-current double-closed-loop feedback technology to improve power supply regulation and load regulation, resulting in high voltage stability and low ripple. Use imported high-power IGBT devices and their driving technology to eliminate switching interference and achieve high quality and portability of Integrated dc high voltage generator. The imported high-frequency high-voltage rectifier diode is used to make the double-press cylinder more compact and lighter, and improve the efficiency of the whole machine. According to EMI electromagnetic interference and EMC electromagnetic compatibility theory, measures such as shielding, isolation and grounding are adopted to improve the safety of the whole machine and to withstand the rated voltage discharge without damage. The instrument has complete protection functions, zero protection, over-voltage protection, over-current protection, and breakdown protection. The protection circuit uses nanosecond-level special sensors, which are fast and reliable, and effectively protect people and equipment. The over-voltage setting uses a digital dial switch, which is easy to operate and has high setting accuracy. Add intelligent high-precision 0.75U function button to facilitate the testing of zinc oxide arrester.

    Integrated dc high voltage generator adopts integrated design scheme: double pressure cylinder and Host are placed in one chassis, combined with new technology, select new device, realize small size, high power, light weight, convenient operation, safe and reliable, power supply output quality The high advantage is the ideal test instrument for the high voltage test industry.

Product Parameters

        Specification         60/2         60/3         60/5         120/2         120/3         120/5
        Rated Voltage (kV)         60         60         60         120         120         120
        Rated Current (mA)         2         3         5         2         3         5
        Rated Power (W)         120         180         300         240         360         600
        Machine Quality (kg)         9         10
        Machine Volume (mm3)         465*390*190         565*390*190
        Double Cylinder Height (mm)         440         535
        Output Voltage Accuracy         ±(1.0% Reading ±2 Words)
        Output Current Accuracy         ±(1.0% Reading ±2 Words)
        Ripple Coefficient         ≤0.5%
        Way of Working         Intermittent use, rated load 30 minutes
        Overload Capability         No-load voltage can exceed 10% of rated voltage for 10 minutes
        The maximum charging current is 1.25 times the rated current
        Power Supply: AC 220V±10% 50Hz
        Working Environment         Temperature: -10~40℃
        Relative humidity: no more than 85% at room temperature of 25℃ (no condensation)
        Altitude: below 1500 meters

Product Features

    ● The output voltage is stable: the high-frequency voltage doubler circuit is applied, and the new PWM pulse width modulation technology and the voltage and current double closed-loop feedback technology are applied to improve the power supply adjustment rate and the load adjustment rate, so that the voltage stability is high and the ripple is small.

    ● Comprehensive protection: full protection function, with zero protection, over-voltage protection, over-current protection, breakdown protection, protection circuit uses nanosecond-level special sensor, action fast and reliable, effectively protect personal and equipment safety.

    ● 0.75U function: Add intelligent high-precision 0.75U function one-button button, press this button, the voltage and current will automatically jump to the 0.75U state, which is conducive to the test of zinc oxide arrester.

    ● Zero boost: The boost potentiometer starts from zero, adopts imported multi-turn potentiometer, the boosting process is stable, and the adjustment precision is high.

    ● Over-voltage setting: over-voltage setting uses digital dialing switch, easy to operate, and has a high setting accuracy.

    ● One-piece design: the choice of integrated design, Host and double pressure cylinder is placed in a chassis. The imported high-frequency high-voltage rectifier diode is used to make the double-pressure cylinder small in size, improve the efficiency of the whole machine, and is easy to carry.

    ● Reliable performance: the key components use high-performance imported components, the pressure roller is coated with special insulating materials, good electrical performance, strong moisture resistance, no leakage.

    ● The operation is simple: the instrument interface each function button, the layout is reasonable, the instructions are clear, easy to learn and use.

Standard Accessories

No. Device name Number
1 Control box 1
2 High-voltage cylinder 1
3 Use's manual book 1
4 Certificate of inspection 1
5 Fuse Some
6 Power cable 1
7 Special grounding wire 1
8 Discharging rod 1
9 Current limiting resistor 1
10 Microampere ammeter 1